Karl Fischer Titrator TITR-40VC

Hanya Tersisa 1 lagi

 


Description

Karl Fischer moisture measurement method has been recognized as the most accurate method by many international standards, such as ISO, ASTM, DIN, BS, and JIS. This method is suitable for the determination of moisture content of various substances. Therefore, Karl Fischer moisture analyzers applying its principle have a wide range of applications for solid, liquid and gas samples.

Features

GENERAL FEATURES

1. LCD display screen.

2. Titration methods, curves and results are displayed in detail.

3. Volumetric KF Titration and Coulometric KF Titration are supported.

4. Data analysis feature helps user review and compare results.

5. Data can be easily transferred to printer via RS-232 communication interface.

6. Reset feature automatically resumes all settings back to factory default options.


Volumetric Karl Fischer Titration

1. Support KF Titration Mode and Titer Detection Mode.

2. Features of auto-filling, auto-purging and auto-mixing of the reagents ensure safe handling of Karl Fischer chemicals.

3. Automatic/manual background drift correction ensures accurate results.

4. Selectable units including mg, mg/L, %, ppm, etc.

5. Settable parameters, including measurement unit, polarization current, stirring rate, titration rate, stop volume, endpoint potential, stop criterion and etc.

6. Store up to 200 titration data sets (GLP-compliant).


Coulometric Karl Fischer Titration

1. Generator cell with diaphragm.

2. Automatic/manual background drift correction ensures accurate results.

3. Selectable units including μg, mg, %, ppm, mg/L, μg/mL, etc.

4. Settable parameters, including measurement unit, polarization current, stirring rate, generator current, endpoint potential, stop criterion and etc.

5. Store up to 200 titration data sets (GLP-compliant).

 

Model

TITR-40VC

Volumetric Titration

Range

(0.1mg~250.0)mg

Unit

mg, mg/L, %, ppm

Polarization Current

(1±0.2)μA, (50±10)μA

Accuracy

Repeatability

≤0.5%

Coulometric Titration

Range

10μg~20mg

Unit

μg, mg, %, ppm, mg/L, μg/mL

Polarization Current

(1±0.2)μA, (50±10)μA

Working Current

for 10.72mA, 20 mA, 53.61mA, 107.2 mA

Working speed

1μg(HO)/s, 1.86μg(HO)/s, 5μg(HO)/s,10μg(HO)/s

Working speed

≤3min for 1000μg water

Accuracy

± (5% of point +3) μg

Repeatability

≤3%

General

Power Supply

AC Adapter, 100-240 V AC input

Dimensions

340*400*400 mm

Net Weight

10kg