X-ray Fluorescence Sulfur Analyzer FSO-X1106 Scitek

Hanya Tersisa 1 lagi

 


Application:

Based on X-ray fluorescence (XRF) spectroscopy, employs high-energy X-rays to directly irradiate oil samples, exciting the inner-shell electrons of sulfur atoms and causing them to emit characteristic X-ray fluorescence. The sulfur content is determined by measuring the energy and intensity of this characteristic X-ray emission. It is commonly used for the routine monitoring of crude oil, heavy oil, residual oil, and refined petroleum products.

 

Features:

1. Windows-based operating platform for user-friendly operation.

2. Non-destructive testing, requires no high-temperature processing or chemical reagents, and allows for sample recovery.

3. Low sample consumption (2–4 mL per test) and rapid detection, providing accurate sulfur content measurements in as little as 30–60 seconds.

4. Uses low-cost, disposable, single-film sample cups to prevent cross-contamination between samples.

5. Sample concentration, ID, and date can be recorded directly on the back of the cup.

6. Rapid instrument calibration using dedicated standard samples, instrument status and electrical parameters are assessed via count and energy spectrum measurements of reference samples.

7. Automatic gain correction, automatically adjusts algorithms to maintain measurement accuracy and extend equipment lifespan if X-ray tube energy output drops due to prolonged use.

8. Rapid instrument calibration is achieved simply by inserting dedicated standard samples.

9. Intelligent assessment of current instrument status and electrical parameters via count and energy spectrum measurements of reference samples

10. Comprehensive spill prevention and multi-layered safety protection: contamination prevention, leak protection, and rigorous X-ray shielding.

11. Massive data storage capacity; analysis results and calibration curve parameters can be queried online at any time, ensuring excellent traceability.

12. Standard compliance: ASTM D4294-16.

 

Model

FSO-X1106

Technology

Energy Dispersive X-ray Fluorescence (EDXRF)

Sample Consumption

per test

Analytical Range

Low Range: 0.0007% to 0.1% (7 mg/kg to 1000 mg/kg / 7 ppm to 1000 ppm);

High Range: 0.1% to 50% (1000 mg/kg to 500000 mg/kg / 1000 ppm to 50 wt%);

Limit of Detection

7 mg/kg (7 ppm)

Measurement Accuracy

At ≈ 20 mg/kg: <± 3 mg/kg (<± 3 ppm);

At ≈ 100 mg/kg: <± 8 mg/kg (<± 8 ppm);

At ≈ 1000 mg/kg: <± 5% (Relative);

At ≈ 1.0%: <± 0.5% (Relative);

Analysis Time

Selectable: 5s, 10s, 15s, 30s, 45s, 60s, 100s, 150s, 200s, 300s, 600s, 1000s

Measurement Replicates

Programmable from 1 to 99 cycles

Calibration Capacity

999 user-defined calibration curves;

999 system-standardized correction curves;

Data Interface

USB, RS-232C

Power Supply

AC 220±10%V, 50/60Hz

Consumption

50W max.

External Size (W×D×H)

320×280×250 mm

Shipping Size (W×D×H)

400×360×330 mm

N.W./G.W.

14/20 kg

Operating Temp. and Humidity

5–40 , ≤85% RH