Elisa Reader MPR-A200HT/MPR-A200

Hanya Tersisa 1 lagi

 

Tags : Elisa Reader  Infitek 

Elisa Reader MPR-A200HT/MPR-A200

Brand INFITEK

Origin CHINA

Description

It’s a high-quality microplate reader based on a monochromator with a wavelength range of 200~1000nm. It can be used for spectral scanning, endpoint method and kinetic detection. Suitable for 96-well plates and 384-well plates with and without lids.

-------------------------------------------------------------------------------------------

Features

High Quality Data and Stable Performance

Double beam optical system has the reference optical channel system, which make the data more stable;

After the instrument is started, the light source, grating, detector, position, etc. are automatically calibrated to ensure stable and reliable operation of the instrument;

Long life xenon lamp which can be used for 109

Choose Detection Wavelength Freely with Raster

It adopts the xenon flash lamp as light source, which chooses the wavelength range from 200-1000nm with 1nm step by grating monochromator for the full spectrum scanning.

---------------------------------------------------------------------------------------------------------------

u-Nano Ultra-Micro Plate (Optional)

Quickly complete high-throughput quantification of micro nucleic acid and proteins without samples dilution

Independent lower computer software, can quickly read the sample concentration and purity report;

1~16 samples can be detected at the same time, only 2~4μL sample volume is needed;

During continuous testing, you only need to wipe off the last batch of samples with dust-free paper.

Printer: Connect the printer through the PC end.

-------------------------------------------------------------------

Powerful and Flexible Software

User authority classification;

Powerful data analysis and process;

Standard curve library;

FTP (File transfer protocol);

Multiple report export types;

Printer: Connect the printer through the PC end.

----------------------------------------------------------------------

Cuvette Mode (only for MPR-A200HT)

Independent cuvette slot;

Detection wavelength 200~1000nm;

With incubation function, RT.+4 °C ~45 °C;

Independent cuvette software can be directly used for endpoint method, kinetics, spectral scanning and standard curve establishment.

 

 -----------------------------------------------------------------------------------------------------------------------------------------------------------

Model

MPR-A200

MPR-A200HT

Cuvette Mode

NO

YES

Microplate Type

96/384 wells

Detection Mode

Absorbance

Light Source

Xenon flash lamp/number of flashes>109

Wavelength Range

200~1000nm

Wavelength Accuracy

2nm

Wavelength Repeatability

0.2nm

Optical System

Monochromator, 1nm step

Reading Range

0~4.0OD

Bandwidth

<2.5nm

Detection System

2 silicon photodetectors, one for measurement, one for reference

Linear @450nm

R2≥0.999, [0.0-3.0 Abs]

Absorbance Accuracy @450nm

± (1.0 % + 0.003 Abs), (0 ~2.0 Abs]; ± 2.0 %, (2.0 ~ 2.5 Abs]

Absorbance Repeatability

@450nm

CV<0.5 % or SD<0.003 accurate mode; CV<1.0% fast mode

Measuring Speed

96-well plate: fast mode <8 seconds, accurate mode <28 seconds (end point method)

Shaking

Linear, 3 speeds adjustable

Temperature Range

RT.+4 °C~45 °C

Temp. Accuracy & Uniformity

±0.5 °C @37 °C, ±0.5 °C @37 °C

User Interface

Built-in software, independent use

Optional

Readerlt-II software, u-Nano Ultra-Micro Plate,

ABS Optical Performance Validation Board

Operation Display

Touch screen input, Android system, 10-inch LCD display full board information, can be connected with keyboard and mouse

Internal Storage

16 G storage, can store more than 20,000 data files

Port

1 type B USB interface, 2 type A USB interfaces, 1 network port

Robotic Arm Compatibility

Temporarily incompatible

Electricity

DC 24 V, 6.67 A

External Dimension(W*D*H)

300*500*260 mm

N.W./G.W.

15.5kg/21kg

Package Dimension(W*D*H)

690*490*500mm